Vicon Vantage V16 Скачать руководство пользователя страница 10

Page 10 of 12

August 2015

Vantage 
Quick Start 
Guide

Configure your system

 3 

For each camera, turn the aperture ring 
(the ring with the f-stops) until the 
aperture is fully open. Later you will turn 
the ring back, increasing the f-stop 
number and closing the aperture.

 4 

In the 

System Resources 

pane or 

Selection

 

editor, select all the cameras by Shift-
clicking. In the 

Properties 

section at the 

bottom of the window (or 

Attributes 

editor)

, change 

Grayscale Mode 

to 

All

 5 

In the 

System Resources

 pane or 

Selection

 

editor, select a camera. For each marker:

a.

In the view pane, zoom in to the center 
of the marker.

     Important

Vantage cameras are supplied pre-focused for 
typical motion capture use, so in most 
situations, no focusing is necessary. The 
following step (b) is applicable only if you need 
to re-focus the camera, otherwise, go straight 
to step c.

b.

If required, turn the camera focus ring 
(shown in the previous illustration)

 

until the 2D image of the marker in the 
view pane is clear and sharp. As you 
rotate the focus ring, note that the 
marker images get smaller and then 
begin to get larger again. Good focus is 
when the marker is at its smallest size.

c.

Observe that, with the aperture fully 
open, the markers appear fully 
saturated, with very little grayscale 
gradation and some merging, as in the 
following image.

d.

Adjust the aperture ring to change the 
amount of light that hits the image 

sensor. Turn the aperture ring to close 
the aperture and obtain good, clear 
images of markers even when they are 
close together. 

If you turn the aperture ring too far, 
until the aperture is almost closed, the 
markers appear very small, lack any 
bright grayscale pixels and the circle-
fitting is poor, as shown below.

Adjust the aperture ring so that the 
markers are not too small, and the 
center is just off-white (not fully 
saturated) when viewed from the 
middle of the volume.

 6 

Repeat Step 5 for each camera.

 7 

When you are happy with the focus and 
aperture, to prevent the rings from moving 
if the camera is knocked, tighten the 
locking screws.

     Caution

To avoid damaging the rings, do not over-
tighten the locking screws.

 8 

Select all of the cameras and change the 
Grayscale Mode property back to Auto.

Содержание Vantage V16

Страница 1: ...Quick Start Guide...

Страница 2: ...tion Systems Limited reserves the right to make changes to information in this document without notice Companies names and data used in examples are fictitious unless otherwise noted No part of this p...

Страница 3: ...h power strobes on board processing Gigabit Ethernet connectivity and a new design Available in high performance resolutions up to 16 megapixels at 120 fps full frame capture speeds PoE switch The sup...

Страница 4: ...cables Camera cables include a ferrite at the camera end only If you are using one or more Vicon Lock units or MX T Series cameras with a Giganet additional cables are supplied Vantage calibration de...

Страница 5: ...r to include height as well as width and depth in your volume calculations Install and connect the cameras 1 Installeachcameraonatripod onthewall or on a ceiling mounted rig You can order mounting bra...

Страница 6: ...PC open Windows Network Connections right click the NIC to which the system is to be connected and then click Properties If you are connecting more than one card right click the first NIC 3 In the Pr...

Страница 7: ...to dongles 3 If you are prompted to restart your PC do so as soon as possible License the Vicon application software Vicon application software uses Safenet licensing and a built in licensing utility...

Страница 8: ...ade the Selection editor make sure all of the cameras in your system are listed 4 Ensure that at the top left of the screen thesystemisshownasliveandconnected Position the cameras 1 In the System Reso...

Страница 9: ...volume Zoom in to a marker by right clicking and draggingthemouse Noticehowthepixels change as you lower the light to view pictures from a grayscale blob Grayscale Mode must be set to All 7 Notice th...

Страница 10: ...otate the focus ring note that the marker images get smaller and then begin to get larger again Good focus is when the marker is at its smallest size c Observe that with the aperture fully open the ma...

Страница 11: ...the system You calibrate the system to enable the cameras to capture subject movements throughout the volume You can also set the volume origin and orientation of axes in relation to the capture volu...

Страница 12: ...old is not too high a good default is 0 2 When I tap a calibrated camera to select it the camera indicates that it has been bumped To change the sensitivity of the camera to taps in the camera s Prope...

Отзывы: