Product Specifications

Type: 

No Touch Thermometer (VNT200) 

Measuring range: 

Body: 34.0 °C – 43.0 °C (93.2 °F – 109.4 °F) 

 

Object: 0 °C – 100.0 °C (32.0 °F – 212.0 °F) 

Resolution:  

0.1 °C (0.1 °F)  

Laboratory Accuracy:    

Body Mode: 

 

±0.2 °C for the range 35.0 °C – 42.0 °C 

 

(±0.4 °F for 95.0 °F – 107.6 °F)

 

±0.3 °C (±0.5 °F) for the range outside  

 

 

35.0 °C – 42.0 °C (95.0 °F – 107.6 °F)

 

Surface Mode: 

 

±2.0 °C for the range 0.0 °C – 100.0 °C 

 

(±3.6 °F for 32.0 °F – 212.0 °F)

This thermometer displays a calculated oral equivalent estimate.
Display:  

Liquid Crystal Display, 4 digits plus special  

 icons 

Acoustic: 

Normal: Green backlight, One long beep for  

 

normal temperature 

 

Elevated or high temperature: Yellow or Red  

 

backlight, 4 double beeps 

 

Erase memory: one long beep 

Operating temperature: 

15 °C – 40 °C (59.0 °F to 104 °F)  

Automatic Switch-off :  

Approx. 20 seconds after any button is pressed

Weight:  

85.6 g (with batteries), 63.0 g (w/o batteries)

Long term storage ranges 

Storage/transport temperature:  

-25 °C to 55 °C (-13 °F – 131 °F) 

Humidity:  

15–95% non-condensing 

Battery:  

(2) AAA Batteries - at least 1000 

measurements 

Pressure:  

700-1060 hPA (0.7-1.06 atm)

This infrared thermometer meets requirements established in ASTM Standard E 

1965-98 (for the thermometer system). Full responsibility for the conformance of 

the product to the standard is assumed by Kaz, USA, Inc., a Helen of Troy Company, 

Health & Home, 400 Donald Lynch Blvd., Suite 300, Marlborough, MA 01752.
ASTM laboratory accuracy requirements for the thermometer only in the display 

range of 36 °C to 39°C (96.8 °F to 102.2 °F) for infrared thermometers is ±0.3 

°C (±0.5 °F), whereas for mercury-in-glass and electronic thermometers, the 

requirement per ASTM Standards E 667-86 and E 1112-86 is ±0.1 °C (±0.2 °F).

This device conforms to the following standards:

•  EN 60601-1:2014 – Medical Electrical Equipment – Part 1: General requirements 

for basic safety and essential performance

•  ISO 80601-2-56:2017 – Medical electrical equipment – Part 2-56: Particular 

requirements for basic safety and essential performance of clinical thermometers 

for body temperature measurement

•  ASTM E1965-98:2016 – Standard Specification for Infrared Thermometers for 

Intermittent Determination of Patient Temperature

•  IEC 60601-1-2:2014 Medical electrical equipment – Part 1-2: General requirements 

for basic safety and essential performance – Collateral Standard: Electromagnetic 

disturbances – Requirement and tests

 

NOTE:

 

DO NOT

 use this device in the presence of electromagnetic or other 

interference outside the normal range specified in IEC 60601-1-2:2014.

•  ISO 15223-1:2016 - Medical devices - Symbols to be used with medical device 

labels, labeling and information to be supplied - Part 1: General requirements

•  EN 1041:2013 Information supplied by the manufacturer of medical devices
•  IEC 60601-1-11:2015 Medical electrical equipment – Part 1-11: General 

requirements for basic safety and essential performance  

–  Collateral standard: Requirements for medical electrical equipment and medical 

electrical systems used in the home healthcare environment

To protect the environment, dispose of empty batteries at appropriate 

collection sites according to national or local regulations. 

Keep out of reach of young children, elderly and pets.

BF Type applied part.

See Instruction of use.

  Protected against solid foreign objects of 12.5 mm diameter and larger. 

Protected against water drops falling vertically when the device is tilted  

up to 15°.

Helen of Troy cares about the environment, please recycle.

Содержание A002372R2

Страница 1: ...VNT200US NoTouch 3 in 1 Thermometer...

Страница 2: ...Product Description NO TOUCH THERMOMETER 2 Instruction Manual 2 3 4 1 5 1...

Страница 3: ...ian Temperature elevation may signal a serious illness especially in neonates and infants or in adults who are old frail or have a weakened immune system Please seek professional advice immediately wh...

Страница 4: ...emperature and RED for a high temperature 4 double beeps indicate temperatures above 37 4 C 99 4 F to alert the user that the individual may have a temperature Color Range Reading Meaning Green 35 8 3...

Страница 5: ...d after pressing the MEMO button To view the next most recent temperature taken press the MEMO button again To clear the memory of previous temperatures press and hold the MEMO button for 4 seconds Ch...

Страница 6: ...other cleaning liquids NEVER insert a sharp object into the scanner area or any other open surface on the thermometer 1YEAR LIMITEDWARRANTY You should first read all Instructions before attempting to...

Страница 7: ...STM Standards E 667 86 and E 1112 86 is 0 1 C 0 2 F This device conforms to the following standards EN 60601 1 2014 Medical Electrical Equipment Part 1 General requirements for basic safety and essent...

Страница 8: ...sit www VicksThermometers com Please be sure to specify the model number 2020 All rights reserved Kaz USA Inc a Helen of Troy Company 400 Donald Lynch Blvd Suite 300 Marlborough MA 01752 Certain trade...

Страница 9: ...nt 12 50 Uncoated Revision 2 21APR20 Release Date Date 20APR20 Scale 1 1 Rerelease Date Colors Special Instructions Dielines Do not print Cyan 0 Magenta 0 Yellow 0 Black 100 Helen of Troy Creative Ser...

Отзывы: