8
Controlling and adjustment functions ALE-NC series operation manual
-
70
-
(1) List of weights used for the span adjustment by model (Unit: g).
Model
ALE322NC
ALE1501NC
ALE8200NC
Selectable
weight on
the menu
320
1500
8200
300
1000
8000
200
1000
5000
100
500
2000
50
200
1000
5
20
100
set
1
to 320
1
to 1500
1
to 8200
(2) The span adjustment by the use of a weight less than the weighing capacity may possibly indicate
<UC> on the display. When this is the case, the weighing accuracy is not guaranteed.
Conditions under which <UC> is indicated when a
sample that is more than two times heavier than the
weight that was used for the span adjustment is
weighed.
8-4-1(2)
Span test with external weight
Make sure to use the external weight which is equal to the weighing capacity of each model.
1
Select the span test with external weight.
Press [Menu] key, then press [F1]-[F4]
keys to go to <632 EX SPAN TEST>.
Press [F4] key to execute.
2
Zero-point adjustment starts.
Display changes to the order of <on 0>
→
“blinking of <on 0>”.
On completion of the zero-point
adjustment, the display automatically
changes to <on FS>.
3
The span test starts.
Place the weight on the center of the
weighing pan.
Display changes to the order of <on FS>
→
“blinking of <on FS>”.
Start of the span test.
On completion of the span test, the display
automatically changes to
< INSTRUMENTAL ER> and the
instrumental error of the balance is
displayed.
Press [F4] key.
<632 EX SPAN TEST> is displayed.
Unload the weight from the weighing pan.
4
Exit the setting menu.
Press [Menu] key to shift to the
measuring mode.