ALE-NC series operation manual 3
Functions related to the operation
-
31
-
3-6
Statistics mode
The statistical operation function collects weight data and indicates maximum, average, and other
statistical values.
This mode is not available on verified balance.
(1) Only “mg” or “g” can be used.
(2) Up to 999 weight data can be saved.
(3) The output timing is fixed to “Output once immediately after [Output] key is pressed” or “Output
once when balance reached stable after [Output] key is pressed”, regardless of the setting
value of <413/423 CONDITION> of “6 External input/output function”.
The setting of <17 WT STABLE>
The output condition
<<ON>>
Once at stable after [Output] key is pressed
<<OFF>>
Once immediately after [Output] key is pressed
1
Select the statistics mode.
Press [Menu] key, then press
[F1]-[F4] keys to go to <11 MODE>.
Press [F4] key to change the
setting value.
Press [F4] key to select.
<<STAT>>: Statistics mode
Press [F4] key to fix.
2
Exit the setting menu.
Press [Menu] key to shift to the
statistics mode.
3
Choose whether or not clear all the data.
Press [F3]/[F4] key to select
whether or not clear all the data.
When there is no data stored, this step
is skipped.
<<YES>>
:
Clear
<<NO>>
:
Not clear
When <<NO>> is selected, weighing
step of the next statistics data starts.
4
Store weighing data.
Place the sample in the weighing
pan.
Press [Output] key to store the
sample weight.
Weighing data is collected and then
output.