51
7. Setup of various data
Command Main Body
C1 C2
Code
Description
Value
Response
I
A
49H
41H
Interval time setup
Interval time
A00: Executed
E01: Command error
* Mark off the interval time between the command, hour, minute, second with “,”; e.g., IA, hh, mm,
ss.
8. Span adjustment and test command
Command Main Body
C1 C2
Code
Description
Value
Response
C 0 43H 30H
Command input
ineffective
C 1 43H 31H
Span adjustment with
built-in weight
(Auto-span adjustment)
C 2 43H 32H
Span test with built-in
weight
C 3 43H 33H
Span adjustment with
external weight
C 4 43H 34H
Span test with external
weight
None
A00: Executed
E01: Command error
* Do not use this command on models without the built-in weight.
9. Actual samples of commands
(1) T
Δ
(CR) (LF)
Execute tare range operation (
Δ
: Space).
(2) O1 (CR) (LF)
Set the balance to continuous output.
(3) O0 (CR) (LF)
Stop the balance output.
(4) O8 (CR) (LF)
Output data from the balance.
(5) IA, 12, 34, 56 (CR) (LF)
Set the interval time to 12 hrs.34 min. 56 sec.
(6) DD (CR) (LF)
Output the date.
(7) DT (CR) (LF)
Output the time.
(8) C1 (CR) (LF)
Execute span adjustment with the internal weight.
*
*