10. Functions of Function 2
10.1 Description of Function 2
Item Set
Value
Description
Disable function
Setup of ID No.
Enable
Disable function
Setup of
Instrumental Error
of Weight in Use
Enable
Disable function
Calibration of
Built-in Weight*
Enable
Do not use.
Use of Weight
Error
Use the weight error specified in the span
adjustment/span test with an external weight.
The denotes a factory-setting.
* Not displayed on models without built-in weight.
10.2 Setup of ID No.
The ID No. is used to conduct printing in compliance with ISO/GLP/GMP. When printing is
required, specify the necessary settings and check the current settings. When the ID No. is
displayed, [
] lights up at the top left of the display, and a maximum of 6 digits can be set. [
]
denotes a blank space. The characters that can be used for this setting are: [
] (Blank space),
[
~
,
~
], and [
].
(1) Keep pressing the Function key while
holding down the Zero/Tare key. When
[
] appears, release the key. The
display changes to [
].
(1)
/
+
Key released
g
↓
(2)
/
(3)
(4)
/
(2) Set the rightmost digit to [
] by pressing the
Zero/Tare key.
(3) Press the Function key. The display changes
from [
] to the ID No. indication
*1
.
(4) Press the Zero/Tare key to begin setting or
changing the ID No. The leftmost digit flashes first.
The flashing cursor denotes the digit that can be
altered.
(5) Change the digit indicated by the flashing cursor.
Press the Zero/Tare key to alter the digit
indicated by the flashing cursor. Each press
increases the number by one.
(5)
/