19
(3) Check of capacity point
When
[
] is displayed, load the span test
weight
*1
on the measurement pan. The display
changes to flash, and the balance automatically
checks the capacity point.
Weight loaded
CAL
CAL
(4) Display of span test result
When the capacity point check is completed, the
balance displays [
] and indicates the result
(error) of the span test in a weight value.
CAL
↓
CAL
g
(5) End of span test
After displaying the span test result, the balance
now shows [
] and returns to measurement
mode. Once you are back in measurement mode,
unload the weight.
CAL
○
g
*1 Use a span test weight of Class E
2
(or equivalent) or better under OIML Standard.
Cautions:
1. While the span test is in progress, do not touch the measurement pan and protect the
balance from shock or vibration. If the balance is moved, the processing may be hated at
the stage of the flashing [
] or [
].
2. To cancel the adjustment, press the Print key. The balance returns to measurement
mode after briefly displaying [
].
3. Avoid conducting the span test immediately after the balance has been powered up. Always
wait for at least 30 minutes before conducting the test.