Chapter 4
Configuring System Settings
Managing Options
22130634
CX300 ComXpert Operation Manual
Rev. 001
4.10.2 Reset Factory Default Settings
The following procedure resets the test set to factory default settings.
1.
Proceed to the
System Home screen
.
2.
Select
System Info
.
3.
Select
Reset Instrument to defaults
.
4.
At prompt, reboot the test set to apply the changes.
4.10.3 Erase Test Set Data
This procedure erases all data stored on the test set.
1.
Proceed and arrange to the
System Home screen
.
2.
Select
System Info
.
3.
Select
Secure erase instrument.
4.11 Managing Options
The following procedures describe how to enable, review, and purchase software options.
4.11.1 Enabling Software Options
Software options may be available on the unit or purchased separately. For information
about purchasing software options, contact the VIAVI Solutions Sales Office for your
region.
See section 3.1.15, “System Info Screen”, on page 3-23
for details about option
information provided on the System Info screen.
NOTE
Remote access default setting is “OFF”; remote access to the test set will be lost when
default settings are restored, and until the settings are configured locally.
CAUTION
When
Secure erase instrument
is performed, the CX300 is not operational until
system software is reinstalled in the test set.