Chapter 5
Test and Measurement Functions
Channel Analyzer
22130634
CX300 ComXpert Operation Manual
Rev. 001
Normal
When
Normal
is selected, the Channel Analyzer displays
the maximum value and the minimum value of the sample
data, coinciding with odd versus even numbered data points.
•
Maximum value (+peak): even-numbered data
point
•
Minimum value (-peak): odd-numbered data point
Mean
When
Mean
is selected, the Channel Analyzer displays the
power level corresponding to the data point of the
corresponding time interval. A sample detector provides a
sample for each trace point on the plot field; each trace point
represents a single sample evenly spaced across the Span
of the frequency domain.
A sample detector is effective for measuring noise-like
signals or low power continuous wave (CW) signals that are
near the analyzer’s noise floor.
NOTE
A sample detector can show an inaccurate reading for the amplitude of a CW
signal if the resolution bandwidth (RBW) is set too narrow (reading will be too
low).
Trace Clear All
Pressing this button clears all traces from the plot field.
Table 5-34 Channel Analyzer Controls and Settings (Continued)
Control/Setting
Description