Chapter 5
Test and Measurement Functions
Spectrum Analyzer
Rev. 001
CX300 ComXpert Operation Manual
22130634
5 - 66
February, 2022
Detectors
The Spectrum Analyzer uses detectors in order to accurately
map the correct signal power to the correct frequency point on
the display. The Detector Type selects the type of math that is
used to identify specific information in a received signal.
The Spectrum Analyzer provides the following detector
selections:
Normal
This setting displays random noise better than the peak
without missing signals.
Peak
This setting displays the highest value in each data point.
Negative Peak
This setting displays lowest value in each data point.
Sample
This setting displays the center value in each data point.
Average Count
This field sets the number of measurements used to calculate
average trace.
Trace Math
The trace math shows the difference of Trace 1 and Trace 2
measurement results. If Trace 1 and Trace 2 are active, the
menu, T1 – T2 -> T5 or T2 – T1 -> T6 become available and
you can perform trace math. When performing the trace math,
the scale is automatically set and the display of scale for the
trace math is on the right side of the screen as below.
T1 - T2 > T5
If
Trace 1
and
Trace 2
are active, this menu is activated.
T2 - T1 > T6
If
Trace 1
and
Trace 2
are active, this menu is activated.
Trace Clear All
This button clears all plot traces.
Table 5-39 Spectrum Analyzer Trace Controls and Settings (Continued)
Control/Setting
Description