Performing Calibration
24
Below figure illustrates a connection setup for Thru calibration only. Prior to this Thru calibration, you must
perform O-S-L calibration as described in the previous pages.
Thru calibration
Procedure
1. Connect the test cable to the
Reflection / RF Out
port and
RF In
port of the instrument.
2. Tap
Thru Cal
to start calibration. A progress bar is displayed to show the calibration progress in
percentage.
3. Check the screen to see if the calibration status is now changed to
Cal ON
.