Functionality
Data export, data review, waveform replay, trend review, data sharing
* : External ECG cable, bandwidth mode set to wide
**: Heart rate is calculated based on average of every 5 to 30 QRS complex.
***: Only for measurement with external ECG cable, bandwidth mode set to wide
8.
Electromagnetic Compatibility
The device meets the requirements of EN 60601-1-2. All the accessories also meet the
requirements of EN 60601-1-2 when in use with this device.
Warnings and Cautionary Advices
Using accessories other than those specified in this manual may result in
increased electromagnetic emission or decreased electromagnetic immunity of
the equipment.
The device or its components should not be used adjacent to or stacked with
other equipment.
The device needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided below.
Other devices may interfere with this device even though they meet the
requirements of CISPR.
When the inputted signal is below the minimum amplitude provided in technical
specifications, erroneous measurements could result.
Portable and mobile communication equipment may affect the performance of
this device.
Other devices that have RF transmitter or source may affect this device (e.g. cell
phones, PDAs, and PCs with wireless function).
Guidance and Declaration - Electromagnetic Emissions
The Health Monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that it is used in such an environment.
Emission tests
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage Fluctuations / Flicker
Emissions IEC 61000-3-3
Complies
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the electromagnetic environment specified below. The customer or the
user of the Health Monitor should assure that it is used in such an environment.
Immunity test
IEC60601 test level
Compliance level
Electromagnetic