Battery run time
Only daily check: > 1000 times
Continuous sleep monitoring: > 12 hours
Pure standby calendar mode: > 3 months
Charge time
Less than 2 hours to 90%
ECG
Lead type
Integrated ECG electrodes
External ECG cable and electrodes
Lead set
Lead I, lead II
Measurement mode
Episode, continuous
Sampling rate
500 Hz
Sampling accuracy
16 bit
Display Gain
1.25 mm/mV, 2.5 mm/mV, 5 mm/mV
10 mm/mV, 20 mm/mV
Sweep speed
25 mm/s
Bandwidth*
0.05 to 40 Hz
Electrode offset
potential tolerance
±300 mV
HR measurement
range
30 to 250 bpm
Accuracy
±2 bpm or ±2%, whichever is greater
ST measurement
range
-0.5 to +0.5 mV
Measurement
summary
Heart rate**, QRS duration, ST segment***,QT/QTc Rhythm analysis
(Regular ECG Rhythm, High Heart Rate, Low Heart Rate, High QRS
Value, High ST Value***, Low ST Value***, Irregular ECG Rhythm,
Unable to analyze)
SpO
2
Standards
Meet standards of ISO 80601-2-61
Measurement accuracy verification: The SpO
2
accuracy has been verified in human experiments
by comparing with arterial blood sample reference measured with a CO-oximeter. Pulse
oximeter measurement are statistically distributed and about two-thirds of the measurements
are expected to come within the specified accuracy range compared to CO-oximeter
measurements.
SpO
2
range
70% to 100%
SpO
2
Accuracy (Arms)
80-100%:±2%, 70-79%:±3%
PR range
30 to 250 bpm
PR accuracy
±2 bpm or ±2%
,
whichever is greater
PI range
0.5-15
Measurement
summary
SpO
2
, PR, PI, Summary (Normal Blood Oxygen, Low Blood Oxygen,
Unable to analyze)
22