Vetek Weighing AB
www.vetek.com
JWL Service Manual
May not be reproduced in any form without written permission
PAGE
16
5. Linear Calibration
☆
☆
☆
☆
If the linear calibration is made, there is no need to make the single
point calibration. If the sensor has a good linearity, make the single
point calibration directly.
(1).Press
to turn on the
unit and release the key when
appears on the LCD, the
linear calibration mode is entered.
(2).Press
to do the zero
point calibration and wait for ON 0
to appear (No weight is put on the
pan at this time).
(3). Wait till ON 1 flickers on the
LCD and put on the weight 1/3 of
the full load. For example, 1/3 of
the weighing 30kg is 10kg.