Version V2C22021
Vesticam
™
User Manual
Page | 20
SETTING UP THE BINOCULAR IR GOGGLES ON YOUR PATIENT
1.
Start with the camera cylinders
off…not attached to the goggles
2.
Follow your normal patient hygiene pathways when applying the goggles. Use effective alcohol/anti-
microbial wipes to clean goggles face plate, silicone eye cups and head band before and after each patient.
Do not use on patients with infective disorders. (see also
3.1.2 “Cleaning and avoiding contamination” and
Appendix 1 full instruction manual)
3.
The goggles are comfortable, high quality and flexible enough to fit most face shapes and sizes. Loosen the
head strap by squeezing the quick release buttons and the goggles flange together
4.
Help your patient to position the goggles for comfort. Most patients can do this themselves, like swimming
goggles.
5.
To adjust for interpupillary distance, loosen the locking rings at the front of the mounting plate by turning
anti-clockwise. Allow the mounting plates
to slide into a comfortable position on your patient’s face. Your
patient should be able to see out of both viewing holes.
6.
Pull rubber tags at each side of goggles to tighten gently, and then lock the mounting plates into place by
firmly tightening the locking rings in a clockwise direction.
7.
Hold the first (right) camera cylinder
so that the lead is exiting the cylinder at “5 o’clock”. Insert the camera
cylinder into the mounting plate, and rotate it cloc
kwise until the camera cylinder “clicks” and locks into
place (lead should be at “6 o’clock” when locked).
8.
If you cannot see your patient’s eye completely on the viewing screen, simply loosen the locking ring, realign
the camera, and retighten the locking ring.
9.
TO FOCUS and fine tune eye position use one hand to stabilize the camera cylinder and the other hand to
move the camera joystick up / down and side / side until the eye image is central on screen. Firmly pull
camera joystick in / out to focus the image of the iris and edge of the pupil. The camera cannot be pushed
into the eye as there is an end stop. It is important to focus the eye to ensure that possible torsion is visible.
10.
Repeat the instructions above (7 -9) to position the second (left) camera.
TO ALLOW FIXATION:
11.
Press the small button at the end of the joystick to activate the fixation light. When the button is IN, the
fixation light is ON. When the button is OUT, the fixation light is OFF.
NOTE: the fixation light will only work when the camera is plugged in to the external recording device
(phone, tablet or computer).
12.
If you wish to use “room view”, position the room camera (optional extra) or computer external camera
to
view your patient.
Vesticam2 goggles are designed to be light tight. Adjust the goggles with your patient so they are comfortable,
secure, light tight, and firm.
If you have further questions or concerns please check www.vesticam.com/support or email