7
Installation
CAUTIONS
IN INSTALLATION
REMEMBER
Use only the included power supply adapter. Using other adapters may
result in fire, electric shock and static electricity removal troubles.
REMEMBER
Install the F120S-E securely. The F120S-E can be damaged or malfunction
if it is dropped.
REMEMBER
Install the F120S-E based on environments stated below.
Adverse environments cause product deterioration and breakdown.
Humidity: 35~65%RH (No dewing)
Temperature: 5~40
℃
PROHIBITED
Do not use the F120S-E based on environments stated below. Adverse
environments can result in explosion, fire, electric shock and accident.
*Where explosion-proofing is required;
*Where flammable atmospheres are found;
*Where a significant vibration and/or impact is given;
*Where temperature and humidity exceed specified ranges;
*Where dewing or sharp temperature fluctuations occurs;
*Where there is heavy dust or smoke;
*Where water, oil, chemicals, etc are used;
*Where strong electric or magnetic fields are generated.
When Used on a Desktop
The F120S-E can be used on the desktop.
Do not locate on unstable tables, etc. the F120S-E can be damaged and/or malfunction if it is
dropped.
When Mounted on a Wall or a Structure
・
Select a location and height in consideration of F120S-E's static electricity removal
range. For information regarding the static electricity removal range,
refer to Static Electricity Removal Performance.
・
Use the mounting holes in the bottom of the stand to secure it to the structure or
cell base. For the dimension drawing, refer to External View.
・
The angle can be adjusted using the adjustment screws located on the left and
right sides of the stand (hexagon socket head bolt, width across flats 4mm).
Securely tighten the adjustment screws after adjustment.
・
Keep objects from getting in between the F120S-E and the target area.
・
Before installing the F120S-E, check the strength of supporting structure.
・
Please keep backside of the F120S-E away from the object and keep
away from the wall surface at least 100mm for the proper function of the fan.
・
A screw is supplied with the F120S-E to ensure more firm fixing of
Front Cover on the body. Use it when needed.
Содержание Stat Clean F120S-E
Страница 5: ...5 External View No F120S E No AD 24 IT EX ...