3.0 301 Controller
3.8.2
Single Point Calibration -
C.CL1
Single point calibration uses an offset value to adjust the temperature over its whole
range.
Single point calibration accurately sets the temperature for setpoint values close to the
temperature at which the calibration offset is made. The accuracy is reduced for
setpoint temperatures which are significantly higher or lower then the calibration offset.
Table showing examples of how to determine offset values:
Measured Calibration
Temp (°C)
Displayed
Temp (°C)
Old Offset
Value (°C)
New Offset
Adjustment
New Offset
Value (°C)
252
250
0
2
2
248
250
0
-2
-2
252
250
2
2
4
The calibration temperature may be measured at the centre of the chamber or through
a specially fitted port.
New Offset Value = Old Offset Value + New Offset adjustment
New Offset Adjustment = Measured Calibration Temperature – Displayed Temperature
Caution! - The procedure to determine the calibration temperature at the
centre of a chamber is not covered in these instructions. If you are unsure
how to do this safely, then seek advice as there is a risk of electric shock if
done incorrectly.
3.8.3
Changing the Single Point Calibration Offset -
OFSt
l
Start at the home display.
l
Hold the Page Key for 1.5 seconds to access the Setup Menu.
l
Repeatedly press the Page key until
CL.St
is displayed.
l
Press the up or down Arrow keys to display the current cal-
ibration offset.
l
If a password is required to access
C.CL1
, enter the password
using the up and down Arrow keys and press the Page key to
accept. You will be returned to
CL.St
.
l
Use the up and down Arrow keys to scroll to
C.CL1
.
l
Press the Page key to access
OFSt
.
l
Press the up or down Arrow keys to change the offset value.
l
The value will then be stored without any further input.
Once
C.CL1
has been selected as the customer calibration type, it is possible to start
from the home display and go to
OFSt
directly, and enter the password at this point
when the calibration adjustment is required again.
3.8.4
Dual Point Calibration -
C.CL2
Dual point calibration uses two offset values at two corresponding temperatures to
progressively change the calibration as the temperature increases or decreases. This is
a more accurate representation of how the temperature difference will occur.
30