59
Operations & Maintenance Manual: Maintenance & Safety
0900‑5005 REV‑00
PROCEDURE 4. COMPLETE A CALISCAN TEST
The BladderScan i10 system includes the
CaliScan
test, which tests the probe’s mechanical components and
transducer function. This test can be used for preventative maintenance or as a diagnostic tool if a probe has
been dropped or in storage for an extended amount of time.
Note: The CaliScan test evaluates only the mechanical condition of the probe. The BladderScan i10 system
does not require calibration as normal scheduled maintenance.
1. Make sure that the BladderScan i10 system is disconnected from external power.
2. On the Home screen, tap
Settings
, and then tap
Self‑Test
.
3. Swipe upward in the list of test results until the Probe section appears.
4. Tap
Start CaliScan™ Test
.
5. Review the test instructions, and then tap
Test
. The system begins to test probe components. A message
and process indicator appear while the test is in process.
Note: Once the test has started, do not press the On/Off button or disconnect the probe from the
console.
6. Proceed according to the message that appears on the screen.
If the message reports that …
Then …
The probe passed the test
When you are done reviewing the results, tap
Back
.
The probe failed the test
Restart the system and try the test again. If the test continues
to fail, contact Verathon Customer Care or your local
representative.