Care and Maintenance
7-16
NeoGraft 2.0® User Manual
7.5. Cleaning the Implantation
Handpiece Components
7.5.1.
Disconnecting the Implantation
Handpiece Components
To disconnect the implantation handpiece components:
1.
Remove the handpiece from the system.
2.
Disconnect the tubes from the handpiece and discard them.
Warning
Used tubes must be discarded as biohazard waste.
3.
Unscrew and disconnect the handpiece head from the
handpiece body.
4.
Unscrew and disconnect the handpiece base from the
handpiece body.
5.
Remove the piston from the handpiece head.
6.
Unscrew and disconnect the stainless steel tube connector
from the handpiece head.