Care and Maintenance
7-10
NeoGraft 2.0® User Manual
To perform automatic cleaning/disinfection of the handpiece
motor:
Use the following guidelines:
•
Verify that the machine washable symbol
is engraved on
the instrument.
•
Use a validated cleaner-disinfectant with the adequate detergent.
•
Select the specified washing cycle for the instrument, which
should be compatible with the detergent manufacturer's
instructions.
•
Use the following recommended temperatures:
▪
Pre-wash – Water: cold to maximum 35°C (95°F) –
Duration 2 min.
Wash – Water: 50°C to 65°C (122°F to 149°F –
Duration 5 min.
▪
Rinsing:
Water: cold to maximum 35°C (95°F) – Duration 2min.
▪
Thermal disinfection:
Water: 80°C to 97°C (176°F to 206.6°F) – Duration 5 min.
Drying:
Air: 65°C to 75°C (149°F to 167°F) – Duration 25 min.