Care and Maintenance
7-8
NeoGraft 2.0® User Manual
10.
Dry the external parts, using a disposable wipe.
11.
Dry the internal parts, using Spraynet.
12.
Visually inspect the parts and make sure that they are clean,
intact and in good condition.
•
Any soiled part must be re-cleaned.
•
Any damaged (e.g. broken, twisted, rusty) part should be
discarded.
13.
Lubricate the contra-angle head with the Lubrifluid.
Caution
In order to avoid corrosion, it is essential to perform proper
lubrication of the component.
14.
Place the contra-angle head vertically on a gauze pad, to
absorb the excess oil, for about 2 minutes.