storage detects and marks these false echoes
,
so
that they are no longer taken into account for the
level measurement
.
A
false echo memory should
be created with empty vessel so that all potential
interfering re
fl
ections will be detected
.
G
ating out of false signals
C
hange now
?
P
roceed as follows
:
1
M
ove from the measured value display to the
main menu by pushing
[
OK
]
.
2
S
elect the menu item
"
S
ervice
"
with
[-
>
]
and
con
fi
rm with
[
OK
]
.
N
ow the menu item
"
F
alse
signal suppression
"
is displayed
.
3
C
on
fi
rm
"
F
alse signal suppression
-
C
hange
now
"
with
[
OK
]
and select in the below menu
"
C
reate new
"
.
E
nter the actual distance from
the sensor to the product surface
.
A
ll false
signals in this area are detected by the sensor
and saved after con
fi
rming with
[
OK
]
.
N
ote
:
C
heck the distance to the product surface
,
because if an incorrect
(
too large
)
value is
entered
,
the existing level will be saved as
false signal
.
T
he
fi
lling level would then no
longer be detectable in this area
.
C
opy sensor data
T
his function enables reading out parameter
adjustment data as well as writing parameter
adjustment data into the sensor via the indicating
and adjustment module
.
A
description of the
function is available in the operating instructions
manual
"
I
ndicating and adjustment module
"
.
C
opy sensor data
C
opy sensor data
?
R
eset
B
asic adjustment
I
f the
"
R
eset
"
is carried out
,
the sensor resets the
values of the following functions to the reset values
(
see chart
):
1
)
F
unction
R
eset value
M
ax
.
adjustment
0
m
(
d
)
M
in
.
adjustment
M
eas
.
range end in m
(
d
)
2
)
M
edium
L
iquid
V
essel form
not known
D
amping
0
s
L
inearization
linear
S
ensor
-
TAG
S
ensor
D
isplayed value
D
istance
C
urrent output
-
characteristics
4
…
20
m
A
C
urrent output
-
max
.
current
20
m
A
C
urrent output
-
min
.
current
4
m
A
C
urrent output
-
failure
<
3
.
6
m
A
U
nit of measurement
m
(
d
)
T
he values of the following functions are
not
reset
to the reset values
(
see chart
)
with
"
R
eset
"
:
F
unction
R
eset value
L
ighting
no reset
L
anguage
no reset
HART
mode
no reset
F
actory setting
L
ike basic setting
,
in addition special parameters
are reset to default values
.
3
)
P
ointer
T
he min
.
and max
.
distance and temperature
values are reset to the actual value
.
O
ptional settings
A
dditional adjustment and diagnosis options such
as e
.
g
.
scaling
,
simulation or trend curve presen
-
tation are shown in the following menu schematic
.
1
)
S
ensor
-
speci
fi
c basic adjustment
.
2
)
D
epending on the senso type
,
see
"
T
echnical data
"
.
3
)
S
pecial parameters are parameters which are set customer
-
speci
fi
cally on the service level with the adjustment
software
PACT
ware
™
.
VEGASON S
61
31
S
et up with the indicating and adjustment module