8
.
5
F
low measurement
F
or
fl
ow measurement in open
fl
umes
,
a constriction or standardised
fl
ume must be used
.
D
epending on the
fl
ow volume
,
this constriction
generates a certain amount of backup
.
T
he
fl
ow can be determined
from the height of this backup
.
T
he
fl
ow volume is outputted by an
appropriate number of pulses on the relay or current output and can
thus be further processed by connected downstream instruments
.
T
here is also the possibility to sum up the
fl
ow volume by means of a
totalizer
,
the result is available on the display and as
PC
/
DCS
value
.
D
epending on the type and version
,
each
fl
ume generates a di
ff
erent
backwater
.
T
he data of the following
fl
umes are available in the
instrument
:
l
P
almer
-
B
owlus
-
F
lume
l
V
enturi
fl
ume
,
trapezoidal weir
,
rectangular weir
l
V
-
N
otch
S
etup
T
he con
fi
guration of the
fl
ow measurement loop requires
PACT
ware
with the suitable
DTM
s
.
T
he example refers to a
fl
ow measurement
with a radar sensor
.
T
he following setup steps must be carried out
:
l
S
election of the parameter
"
F
low
"
l
C
arrying out adjustment
l
S
elect
fl
ume
(
linearization
)
l
S
et scaling
l
P
arameter adjustment of pulse outputs
l
P
arameter adjustment of the totalizer
S
elect in the
DTM
window
"
P
arameter
"
the option
"
F
low
"
with the
requested unit of measurement
.
M
in
.
adjustment
:
E
nter the suitable value for
0
%
,
i
.
e
.
the distance
from the sensor to the medium as long as there is no
fl
ow
.
T
hese are in
the following example
1
.
40
m
.
M
ax
.
adjustment
:
E
nter the suitable value for
100
%
,
i
.
e
.
the distance
from the sensor to the medium
,
with the max
.
fl
ow volume
.
T
his is in
the following example
0
.
80
m
.
F
unctional principle
F
lume
P
arameter
-
F
low
O
peration
48
VEGAMET
391
•
4
…
20
m
A
/
HART
signal conditioning instrument
8
A
pplication examples
36032
-
EN
-
091116