10 Diagnostics and servicing
VEGAFLEX 82 • 4 … 20 mA/HART - two-wire
44222-EN-170220
Code
Text message
Reason
Rectification
DevSpec State
in CMD 48
F105
Measured value
is determined
•
The instrument is still in the start
phase, the measured value could
not yet be determined
•
Wait for the end of the switch-on
phase
•
Duration depending on the version
and parameter adjustment max.
5 min.
Bit 6 of
Byte 0 … 5
F113
Communication
error
•
EMC interference
•
Transmission error during external
communication with 4-wire power
supply unit
•
Remove EMC influences
•
Exchange 4-wire power supply
unit or electronics
Bit 12 of
Byte 0 … 5
F125
Impermissible
electronics tem-
perature
•
Temperature of the electronics in
the non-specified range
•
Check ambient temperature
•
Insulate electronics
•
Use instrument with higher tem-
perature range
Bit 7 of
Byte 0 … 5
F260
Error in the cali-
bration
•
Error in the calibration carried out
in the factory
•
Error in the EEPROM
•
Exchanging the electronics
•
Send instrument for repair
Bit 8 of
Byte 0 … 5
F261
Error in the in-
strument settings
•
Error during setup
•
Error when carrying out a reset
•
False signal suppression faulty
•
Carry out a reset
•
Repeat setup
Bit 9 of
Byte 0 … 5
F264
Installation/Set-
up error
•
Error during setup
•
Check for correct mounting and/or
parameter settings
•
Check probe length
Bit 10 of
Byte 0 … 5
F265
Measurement
function dis-
turbed
•
Sensor no longer carries out a
measurement
•
Carry out a reset
•
Disconnect operating voltage
briefly
Bit 11 of
Byte 0 … 5
F266
Impermissible
operating voltage
•
Operating voltage below specified
range
•
Check electrical connection
•
If necessary, increase operating
voltage
Bit 14 of
Byte 0 … 5
F267
No executable
sensor software
•
Sensor cannot start
•
Exchanging the electronics
•
Send instrument for repair
No communica-
tion possible
The following table shows the error codes and text messages in the
status message "
Function check
" and provides information on causes
as well as corrective measures.
Code
Text message
Reason
Rectification
DevSpec
State in CMD 48
C700
Simulation active
•
A simulation is active
•
Finish simulation
•
Wait for the automatic end after
60 mins.
"Simulation
Active" in "Stand-
ardized Status 0"
C701
Parameter verifi-
cation
•
Parameter verification was inter-
rupted
•
Finish parameter verification
Bit 12 of
Byte 14 … 24
Function check