3 Product description
VEGAFLEX 82 • 4 … 20 mA/HART - two-wire
44222-EN-170220
•
Download the smartphone app "
VEGA Tools
" from the "
Apple App
Store
" or the "
Google Play Store
"
•
Scan the Data Matrix code on the type label of the instrument or
•
Enter the serial number manually in the app
This operating instructions manual applies to the following instrument
versions:
•
Hardware from 1.0.0
•
Software from 1.3.0
•
DTM from version 1.67.2
The instrument and the electronics version can be determined via the
product code on the type label as well as on the electronics.
•
Standard electronics: Type FX80H.-SIL
The scope of delivery encompasses:
•
Sensor
•
Optional accessory
•
Optionally integrated Bluetooth module
•
Documentation
– Quick setup guide VEGAFLEX 82
– Safety Manual (SIL)
– Instructions for optional instrument features
– Ex-specific "
Safety instructions
" (with Ex versions)
– If necessary, further certificates
•
DVD "
DTM Collection
", included therein
– PACTware
– DTM Collection
– Instrument master files (GSD) for Profibus PA
– FDT certificates
Information:
In this operating instructions manual, the optional instrument features
are described. The respective scope of delivery results from the order
specification.
3.2 Principle of operation
The VEGAFLEX 82 is a level sensor with cable or rod probe for con-
tinuous level measurement, suitable for applications in bulk solids.
Due to the qualification up to SIL2 or homogeneous redundant up to
SIL3 (IEC 61508) the VEGAFLEX 82 is suitable for the use in safety-
instrumented systems (SIS).
The safety function (SIF) can be a monitoring of the max. or min. level
or a combination of both.
High frequency microwave pulses are guided along a steel cable or
a rod. Upon reaching the product surface, the microwave pulses are
reflected. The running time is evaluated by the instrument and output-
ted as level.