PaX-i (PCH-2500) User Manual
19
2. Warnings and Precautions
CAUTION
Risk of eye injury!
Do not use this equipment with any other laser sources
and do not make any changes to the settings or processes
that are described in these operating instructions.
Cleaning
■
Never expose this equipment to liquids, mists or sprays. Exposing this
equipment to liquids may cause an electrical shock or otherwise damage
the system.
■
Do not use spray cleaners on this equipment, as this could cause a fire.
During Operation
■
Never use this equipment in an environment that is susceptible to explosion.
■
Do not place flammable materials near this equipment.
■
Do not operate the PC or Touchscreen while the equipment is performing
an operation. Failure to comply with this instruction may result in system
malfunction.
■
Immediately stop imaging if the equipment malfunctions in any way.
■
If a problem occurs during imaging, press the red emergency stop button
located on top of the handle frame to immediately stop all moving parts and
cut off all power to the equipment’s electrical components.
■
Never touch the patient while also touching the SIP/SOP connectors.
In case of an electrical fire
■
Use only fire extinguishers designed for electrical fires to extinguish fires on
this equipment. Liquid extinguishers, such as those which use water, could
damage the equipment and/or cause injury.
■
Unplug the equipment’s power cable before extinguishing any fire.
Pi_u272_en_20180514.indd 19
2018-05-14 오전 11:03:09
Содержание PAX-I
Страница 1: ...User Manual Full version Model PCH 2500 Version 2 72 English...
Страница 2: ......
Страница 38: ...This page is left intentionally blank...
Страница 101: ...Troubleshooting 8...
Страница 103: ...Cleaning and Maintenance 9 9 1 Cleaning 104 9 2 Maintenance 105...
Страница 107: ...Disposing of the Unit 10...
Страница 136: ......
Страница 137: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...