PaX-i (PCH-2500) User Manual
128
Appendices
P o w e r
f r e q u e n c y
(50/60 Hz)
m a g n e t i c
field
IEC 61000-
4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
l o c a t i o n i n a t y p i c a l
commercial or hospital
environment.
NOTE) U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration - electromagnetic immunity
The model PaX-i(PCH-2500) is intended for use in the electromagnetic
environment specified below. The customer or the user of the model PaX-
i(PCH-2500) should assure that it is used in such an electromagnetic
environment
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Conducted
R F I E C
6 1 0 0 0 - 4 - 6
Radiated RF
IEC 61000-
4-3
3 V r m s
150 kHz to
80MHz 3 V
/ m 80 MHz
to 2.5GHz
3 Vrms 150
k H z t o 8 0
MHz 3 V / m
80 MHz to
2.5 GHz
T h e m o d e l P a X -
i ( P C H - 2 5 0 0 ) m u s t b e
used only in a shielded
location with a minimum
RF shielding effectiveness
and, for each cable that
exits the shielded location,
a m i n i m u m R F f i l t e r
attenuation of 20 dB from
30 MHz to 230 MHz, 20 dB
from 230 MHz to 1 GHz.
Field strengths outside
the shielded location from
f i x e d R F t r a n s m i t t e r s ,
a s d e t e r m i n e d b y a n
electromagnetic site survey,
should be less than 3 V /
m.a Interference may occur
in the vicinity of equipment
marked with the following
symbol:
Pi_u272_en_20180514.indd 128
2018-05-14 오전 11:03:33
Содержание PAX-I
Страница 1: ...User Manual Full version Model PCH 2500 Version 2 72 English...
Страница 2: ......
Страница 38: ...This page is left intentionally blank...
Страница 101: ...Troubleshooting 8...
Страница 103: ...Cleaning and Maintenance 9 9 1 Cleaning 104 9 2 Maintenance 105...
Страница 107: ...Disposing of the Unit 10...
Страница 136: ......
Страница 137: ...Postal code 18449 13 Samsung 1 ro 1 gil Hwaseong si Gyeonggi do Korea www vatech co kr...