6. Acquiring PANO Images
WARNING
Be careful not to project the laser beams directly into the
patient’s eyes as this could severely damage the patient’s
vision.
A.
Mid-sagittal plane laser beam:
Position the Mid-sagittal plane laser beam
at the center of the patient’s face to prevent magnifications of the left or right
side in the final image.
B.
Frankfurt plane laser beam:
Position
the patient’s head so that the Frankfurt
plane is aligned with the Frankfurt plane
laser beam.
C.
Canine laser beam:
Position the canine laser beam on the left corner of the
patient’s lips.
Completing Patient Positioning
A. After confirming patient positioning and beam alignment, adjust the temple
supports to fit snugly on both sides of the patient’s head using the temple
support wheel. The temple support wheel is located underneath the patient
support rest.
Adjusting Lever for
the Frankfurt plane
laser
Adjusting lever
for the canine laser
Canine laser beam
Frankfurt plane laser
beam
The mid-sagittal plane
laser beam
Pi_u272_en_20180514.indd 75
2018-05-14 오전 11:03:25