OPERATION
Series 48
16/27
Edition 2013-05-24
288914ED
Action
Solenoid
Procedure
To close
the valve
Without solenoid
48 ... - . E14- ....
48 ... - . E24- ....
48 ... - . E72- ....
At first, supply admissible compressed air to
connection OPEN.
Afterwards, supply admissible compressed air to
connection CLOSE and release air through
connection OPEN.
Standard solenoid
48 ... - . E44- ....
48 ... - . E74- ....
Supply admissible compressed air to connection
OPEN/CLOSE (PORT 1).
Release voltage from connection SV.
Impulse solenoid
(option)
48 ... - . E44-xxxx
48 ... - . E74-xxxx
xxxx = option code
Supply admissible compressed air to connection
OPEN/CLOSE (PORT 1).
Supply a voltage impulse (observe voltage
specification on coil) to connection SV CLOSE.
Impulse must last for at least 50 ms.
Table 5-2
5.2
Operation under increased temperature
Maximum allowed temperature, see product data sheet.
NOTICE
Inconstant temperatures
Performance of the valve may deteriorate.
– Actuate valve only after the bake-out temperature has been stable for two
hours.
– If valve must be actuated during bake-out, make sure that the heating or
cooling rate does not exceed 10 °C per hour in the temperature range
from 100 °C to 200 °C (version with compact actuator).
from 100 °C to 300 °C (version with extended actuator).
– Make sure that the temperature differences over the whole body do not
exceed 30 °C.