INSTRUCTIONS FOR USE
68 87-900-940-01 (
F
)
When a fault has occurred it is possible to use the
Varian repair service. Replacement controllers are
available on an advance exchange basis through
Varian.
!
Before carrying out any work on the controller,
disconnect it from the supply.
If a pump is to be scrapped, it must be disposed of in
accordance with the specific national standards.
ERROR MESSAGES
For a certain type of failure, the controller will self-
diagnose the error and the messages described in
the following table are displayed.
MESSAGE
DESCRIPTION
REPAIR ACTION
CHECK CONNECTION
TO PUMP
Wrong connection between the pump
and the controller.
Check connection between controller and
pump.
Press the START push-button twice to start
the pump.
PUMP WAITING
INTERLOCK
The interlock signal of P1 connector is
activated by an interruption of the link
between pin 3 and 8 of J1 connector, or
because the external interlock signal is
open.
Reset the short circuit between pin 3 and
pin 8 of J1 connector, or close the external
interlock signal.
FAULT: OVERTIME SX
Within each step of the soft start mode
the rotational speed of the pump does
not reach the planned value within 15
minutes.
(X) is the step number from 0 to 9
indicating the step number not passed).
Verify that system has no leaks. Press the
START pushbutton twice to start the pump.
FAULT: PUMP OVERTEMP.
The upper bearing/pump temperature
exceeds 60 °C.
Wait until the temperature decrease below
threshold value.
Press the START push-button twice
to start the pump.
FAULT: CONTROLLER
OVERTEMPERATURE
The controller transformer temperature
exceeds 90 °C.
Wait until the temperature decrease below
threshold value.
Press the START push-button twice
to start the pump.
FAULT: TOO HIGH L0AD
In normal operation (after the starting
phase) the current drawn by the pump is
higher than programmed.
Check that the pump rotor is free to rotate.
Press the START push-button twice to start
the pump.
FAULT: SHORT CIRCUIT
After the starting phase the output
connection is shorted.
Check connections and shortages between
pump and controller.
Press the START push-button twice
to start the pump.
SYSTEM OVERRIDE
The pump is stopped by an emergency
stop signal provided via a remote
contact.
Remove the controller power cable and
check the emergency condition. Then
reconnect the power cable and
press the START push-button twice
to start the pump.
OVERVOLTAGE
Controller power supply circuitry is
faulty, or the Controller received a spike.
Press the START push-button twice to start
the pump.
Should the message still be present, call
the Varian service.
CONTROLLER FAILURE
Fuse blown on the power supply
secondary of the transformer.
Contact Varian for Maintenance.
WARNING!
Содержание Turbo-V 301
Страница 2: ...Turbo V 301 Rack Controller...
Страница 3: ......