TECHNICAL INFORMATION
7
0 87-900-940-01 (
F
)
CONTROLLER SPECIFICATIONS
Input:
Voltage
Frequency
Power
100, 120, 220, 240 Vac
±
10%,
47 to 63 Hz
450 VA maximum
Output:
Voltage
Frequency
Power
75 Vac nominal
±
10%,
3-phase
933 Hz,
±
2%
150 W maximum
Operating
temperature
0 °C to +40 °C
Storage temperature
-20 °C to +70 °C
Fuse (mains)
2 x T4 A (slow blow)
disregarding the mains
P1 optoisolator input
Minimum ON 3 mA
Maximum 5 mA
J2 optoisolator output
24 Vdc, 60 mA
Radio interference
suppression
EN 55011 class A group 1
EN 61010-1
IEC1000-4-2,
1000-4-3, 1000 4-4
P6 output fan
150 mA (24 Vdc)
J7 output
(forepump coil, etc.)
120 Vac, 1 A
disregarding the mains
Interconnecting
cables
Mains cable (3-wire,
3-meter long)
Pump cable (6-wire,
5-meter long)
Weight:
4.5 Kg (8.4 lbs)
Pollution Degree
2
Installation Category
II
CONTROLLER OUTLINE
The outline dimensions for the Turbo-V 301 Rack
controllers are shown in the following figure.
Controller Outline