background image

TECHNICAL INFORMATION

2 87-900-944-01 (C)

DESCRIPTION

The Turbo-V 300 box controller is a microprocessor-
controlled, solid-state, frequency converter with  self-
diagnosis and protection features.

The controller drives the Turbo-V 300 pump series by
controllling the voltage and current respect to the speed
reached by pump.

It incorporates all the facilities required for the operation
of the Turbo-V 300 pump series: pump start/stop, digital
current and speed control, analog signals for external
indicators.

The power is externally supplied.

All the input/output connections are performed on:

  9 pin “D” type male connector attached to a cable

400 mm long for I/O and Electrical DC supply.

  Pump connection with 400 mm long cable.

  9 pin “D” type connector for RS232 connection.

TURBO-V 300  75 VDC BOX CONTROLLER
DESCRIPTION

The controller is a solid-state frequency converter which
is driven by a single chip microcomputer and is
composed of a PCB which includes all the circuitry
necessary for its operation.

The microcomputer generates the variable output
voltage according to the software and the gas load
condition of the pump.

Moreover, it manages signals from sensors,
input/output connection information, and gives output
for a fully automatic operation.

The controller can be operated via remote signals
through an RS-232 connection.

The controller can be operated in local mode through
suitable switches connected between the input pins of
the TV300 connector.

CONTROLLER SPECIFICATIONS

Input:
- Voltage

- Current

75 Vdc
with 2 Vpp max ripple

3.0 A max.

Fuse

T 3 A

Output:
- Voltage

- Frequency
- Power

80 Vac nominal ±10%,
3-phase

933 Hz, ±2%

165 W maximum

Compliance to Norms:

-  Radio interferences

-  ESD

-  BURST

-  Radiated RF
   immunity

-  Safety

EN 55011 Class Group 1

EN 61000/4/2

EN 61000/4/4

EN 61000/4/3

EN 61010/1

Installation category

II

Operating temperature

0°C to + 40 °C

Storage temperature

-20°C to + 70°C

Cooling

Internal fan

Weight 

0.5 Kg (1.1 Ibs)

 

!

There can be 75 Vdc voltage referred to ground on the
pump cable or on the serial connector.

WARNING!

Содержание Turbo-V 300 SQ 275

Страница 1: ...87 900 944 01 C JULY 2006 GB INSTRUCTION MANUAL Turbo V 300 75 Vdc Box Controller Model SQ 275 vacuum technologies...

Страница 2: ...Turbo V 300 75 Vdc Box Controller...

Страница 3: ......

Страница 4: ......

Страница 5: ...ump Connector 4 I O Specifications 4 9 pin D Type Connector Pin Assignement 4 Serial Communication Port 5 RS 232 Communication Descriptions 5 Transmission Channel Characteristics 5 Message Structure 5...

Страница 6: ...azards Warning are used when failure to observe instructions or precautions could result in injury or death CAUTION Cautions are used when failure to observe instructions could result in damage to equ...

Страница 7: ...and is composed of a PCB which includes all the circuitry necessary for its operation The microcomputer generates the variable output voltage according to the software and the gas load condition of th...

Страница 8: ...TECHNICAL INFORMATION 3 87 900 944 01 C CONTROLLER OUTLINE The outline dimension for the controller are shown in the following figure Controller outline...

Страница 9: ...ins A F Pump temperature sensor Pin E Ground I O Specifications START STOP START command STOP command Low 0 8 Vdc High 4 to 15 Vdc Analog output Output impedence Minimum load 0 to 10 Vdc proportional...

Страница 10: ...ommunication RS 232 serial port connections Transmission Channel Characteristics levels RS 232 RS 422 baud rate 9600 4800 programmable by a jumper on the board character length 8 bits parity none stop...

Страница 11: ...35 STX ADDR ACK ETX CRC Command SOFT START ON Source PC Destination Inverter 02 80 31 30 30 31 31 03 42 32 STX ADDR WINDOW WR ON ETX CRC Source Inverter Destination PC 02 80 06 03 38 35 STX ADDR ACK E...

Страница 12: ...4 brake 5 normal 6 failure 206 N X ERROR CODE Too high load No connection Short circuit Pump overtemp SoftStart Not Ended Controller overtemp RunUpTime Not Reached Power fail 211 N X PUMP SENSOR TEMPE...

Страница 13: ...9 pin D type connector To deactivate the Low Speed status it is necessary to disconnect pin 9 from pin 15 ground of the same connector The low speed frequency is equal to 622 Hz MAINTENANCE Replacemen...

Страница 14: ...TECHNICAL INFORMATION 9 87 900 944 01 C PCB JUMPERS...

Страница 15: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 16: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 17: ...1263 Toll Free 0120 655 040 Korea Varian Technologies Korea Ltd Shinsa 2nd Bldg 2F 966 5 Daechi dong Kangnam gu Seoul Korea 135 280 Tel 82 2 3452 2452 Fax 82 2 3452 2451 Toll Free 080 222 2452 Mexico...

Отзывы: