TECHNICAL INFORMATION
7 87-900-944-01 (C)
Serial Communication Windows
WIN
TYPE
R W
Description
000
L
X
X
START/STOP (1= START ; 0= STOP)
008
L
X
X
REMOTE/SERIAL Configuration (1= Remote ; 0= Serial)
100
L
X
X
SOFT START YES/NO (1= YES ; 0= NO)
Default= 0
107
L
X
X
ACTIVE STOP (0=NO; 1=YES)
Default= 1
108
N
X
X
BAUD RATE (3-4) [4800-9600]
Default= 4
109
L
X
PUMP LIFE RESET [Write “1” to Reset]
120
N
X
X
SET ROTATIONAL FREQUENCY [Hz] 150 Hz < = F_imp < = FMAX
121
N
X
X
MAX SETTABLE ROTATIONAL FREQUENCY [Hz] F<=F_MAX_ABS
130
N
X
RAMP CURRENT [mA]
200
N
X
CURRENT [mA]
201
N
X
VOLTAGE[V]
202
N
X
POWER [W]
203
N
X
DRIVING FREQUENCY [Hz]
204
N
X
PUMP TEMPERATURE [°C]
205
N
X
STATUS [0=stop; 1=interlock; 2=ramp; 3=regulation; 4=brake; 5=normal; 6=failure]
206
N
X
ERROR CODE:
Too high load
No connection
Short circuit
Pump overtemp
SoftStart Not Ended
Controller overtemp
RunUpTime Not Reached
Power fail
211
N
X
PUMP SENSOR TEMPERATURE READING [208= 25°C - 128= 60°C]
216
N
X
AMBIENT SENSOR TEMPERATURE READING
300
N
X
CYCLE TIME [min]
301
N
X
CYCLE NUMBER
302
N
X
PUMP LIFE [h]
319
A
X
Controller Model
320
A
X
Base Pump Model Number (8 characters)
321
A
X
Modified Standard Model Number (4 characters)
323
A
X
Controller Serial Number (5 characters)
325
A
X
Electrical Modification Level (10 characters)
400
A
X
CRC PROGRAM LISTING [QE7xxxx]
401
A
X
CRC BOOTLOADER [BL1xxxx]
402
A
X
CRC PARAMETER LISTING [PA7xxxx]
404
A
X
CRC FILE PARAMETER STRUCTURE
406
A
X
PROGRAM LISTING CODE & REVISION
407
A
X
PARAMETER LISTING CODE & REVISION
500
L
X
MONITOR MODE
WIN = Window
L = Logical
R =
Read
N = Numeric
W =
Write
A = Alphanumeric
7