TECHNICAL INFORMATION

5 87-900-944-01 (C)

Serial Communication Port

Communication serial port connections and mini- mum
connection configuration are shown in the following
figures. The communication port mating connector is
supplied with the RS 232 PCB (AMP/Cannon or
equivalent 9-pin "D" type male connector). The external
cable (not supplied) between the host computer and the
controller does not require crossed wires so that signals
are connected correctly .
For example, the Transmit data signal from controller
(pin 2) must be connected to the host computer's
Receive data line (pin 2) and vice versa. Consult the
host computer's instruction manual for its serial port
connections

NOTE

Varian cannot guarantee compliance with FCC
regulations for radiated emissions unless all external
wiring is shielded, with the shield being terminated to
the metal shroud on the O-subconnector. The cable
should be secured to the connector with screws.

 

!

In order to avoid possible conflicts on the Serial Line, it
is advisable to use a 3 wire shielded cable for the TxD,
RxD and GND connections and to leave all the other
pins unconnected.

RS 232 Communication Descriptions

Communication RS 232 serial port connections

Transmission Channel Characteristics

levels: 

RS 232/RS 422

baud rate: 

9600/4800
programmable by a jumper on the 
board

character length:  8 bits
parity: none
stop bit: 

1 bit

protocoll: 

master (PC) / slave (converter)

In this case the value to be assigned to the ADDRESS
field must be 80 hex (for RS 232).

Message Structure

(request and answer have the same format)

The master system (PC) starts every session sending
the following message to the slave units connected:

<STX> / <ADDR>  + <WINDOW> + <COMMAND> +
<DATA> + <ETX> + <CRC>

where:

<STX>=

0x02

<ADDR> =

0x80 (for RS 232 and
RS 422 only)

<ADDR> =

0x80 + device number (0...31)

0xFF: brodcasting command 
(recognized by all the devices, it 
doesn’t implicate any answer)
(for RS 485 only)

<WINDOWS>=

'000'...' 999' window number
the meaning of the window
depends to the device type

<COMMAND>=

0x30 :window value reading
0x31 :window writing

<DATA> =

alphanumeric ASCII string
containing, in the case of writing
operation, the parameter to input
into  the window  addressed by
the field <WINDOW>This field
may have variable length
according to the data type
contained in the window where
you are working in. In the case of
reading request of a window, the
data field doesn’t exist.

<ETX>=

0x03

<CRC>=

XOR among all the characters
following <STX>=(with exception of
<STX>), including the end
character <ETX> hexadecimally
encoded by two ASCII characters.

 

WARNING!

Содержание Turbo-V 300 SQ 275

Страница 1: ...87 900 944 01 C JULY 2006 GB INSTRUCTION MANUAL Turbo V 300 75 Vdc Box Controller Model SQ 275 vacuum technologies...

Страница 2: ...Turbo V 300 75 Vdc Box Controller...

Страница 3: ......

Страница 4: ......

Страница 5: ...ump Connector 4 I O Specifications 4 9 pin D Type Connector Pin Assignement 4 Serial Communication Port 5 RS 232 Communication Descriptions 5 Transmission Channel Characteristics 5 Message Structure 5...

Страница 6: ...azards Warning are used when failure to observe instructions or precautions could result in injury or death CAUTION Cautions are used when failure to observe instructions could result in damage to equ...

Страница 7: ...and is composed of a PCB which includes all the circuitry necessary for its operation The microcomputer generates the variable output voltage according to the software and the gas load condition of th...

Страница 8: ...TECHNICAL INFORMATION 3 87 900 944 01 C CONTROLLER OUTLINE The outline dimension for the controller are shown in the following figure Controller outline...

Страница 9: ...ins A F Pump temperature sensor Pin E Ground I O Specifications START STOP START command STOP command Low 0 8 Vdc High 4 to 15 Vdc Analog output Output impedence Minimum load 0 to 10 Vdc proportional...

Страница 10: ...ommunication RS 232 serial port connections Transmission Channel Characteristics levels RS 232 RS 422 baud rate 9600 4800 programmable by a jumper on the board character length 8 bits parity none stop...

Страница 11: ...35 STX ADDR ACK ETX CRC Command SOFT START ON Source PC Destination Inverter 02 80 31 30 30 31 31 03 42 32 STX ADDR WINDOW WR ON ETX CRC Source Inverter Destination PC 02 80 06 03 38 35 STX ADDR ACK E...

Страница 12: ...4 brake 5 normal 6 failure 206 N X ERROR CODE Too high load No connection Short circuit Pump overtemp SoftStart Not Ended Controller overtemp RunUpTime Not Reached Power fail 211 N X PUMP SENSOR TEMPE...

Страница 13: ...9 pin D type connector To deactivate the Low Speed status it is necessary to disconnect pin 9 from pin 15 ground of the same connector The low speed frequency is equal to 622 Hz MAINTENANCE Replacemen...

Страница 14: ...TECHNICAL INFORMATION 9 87 900 944 01 C PCB JUMPERS...

Страница 15: ...TIFICATION Product Description Varian P N Varian S N Purchase Reference TYPE OF RETURN check appropriate box Paid Exchange Paid Repair Warranty Exchange Warranty Repair Loaner Return Credit Shipping E...

Страница 16: ...rate No zero high backround Gauge tube not working Display problem Vacuum system unstable Cannot reach test mode Communication failure Degas not working Failed to start Other Error code on display Oth...

Страница 17: ...1263 Toll Free 0120 655 040 Korea Varian Technologies Korea Ltd Shinsa 2nd Bldg 2F 966 5 Daechi dong Kangnam gu Seoul Korea 135 280 Tel 82 2 3452 2452 Fax 82 2 3452 2451 Toll Free 080 222 2452 Mexico...

Отзывы: