TECHNICAL INFORMATION
63
87-900-852-01 (B)
•
Enter the value by pressing the PUMP CURRENT
pushbutton, and the display shows:
R S
2
3
2
H O S T
/
P R
I
N T
S E L E C T
I
O N
:
X X X X X
where:
XXXX
== means HOST or PRINT.
•
Select HOST or PRINT by pressing the CYCLE
pushbutton.
R S
2
3
2
H O S T
/
P R
I
N T
S E L E C T
I
O N
:
H O S T
With the RS 232 connected, a bidirectional
communication is established by selecting HOST. Data
are sent to an external computer every time the
external computer asks for the values.
The data available are:
•
Pump/controller operating condition
•
Cycle
time
•
Pump
life
•
Pump
temperature
•
Pump
current
•
Pump
voltage
•
Controller output frequency
•
Cycle
number
•
R1
condition
•
R2
condition
•
Pump life and cycle # zeroing
•
Configuration parameter readings
•
Configuration parameter setting
If PRINT is selected and a printer is connected on RS
232 line, a unidirectional communication is established
and every minute the data are sent to the printer, even
if the pump is not running.
The set of data available are:
•
Pump speed KRPM
•
Pump
temperature
•
Pump current A
•
Pump power W
•
R1
condition
•
R2
condition
The controller is factory-set to HOST.
•
Confirm the selection by pressing the PUMP
CURRENT pushbutton.
•
In this way you enter into an operating phase
named "Monitor Relay Programming" described in
the following paragraph.
Содержание Turbo-V 250 series
Страница 2: ......
Страница 3: ...Turbo V 250 Controllers...
Страница 4: ......
Страница 5: ......
Страница 58: ...TECHNICAL INFORMATION 51 87 900 852 01 B Controller model 969 9425 outline...
Страница 78: ...TECHNICAL INFORMATION 71 87 900 852 01 B R...
Страница 79: ...TECHNICAL INFORMATION 72 87 900 852 01 B...