TECHNICAL INFORMATION
8
4 87-900-913-01 (
N
)
With the RS 232 connected, a bi-directional
communication is established by selecting HOST. Data
are sent to an external computer every time the
external computer asks for the values.
The data available are:
•
High/low speed adjust
•
Pump/controllers operating condition
•
Cycle
time
•
Pump
life
•
Pump
temperature
•
Pump
current
•
Power
•
Pump
voltage
•
Controller output frequency
•
Cycle
number
•
K1
condition
•
K2
condition
•
Life time and cycle # zeroing
•
Configuration parameter readings
•
Configuration parameter setting
Note that the new input value are put in effect only at
the next STOP/START of the pump.
If PRINT is selected and a printer is connected on RS
232 line, an unidirectional communication is
established and every minute the data are sent to the
printer, even if the pump is not running.
The set of data available are:
•
Pump speed KRPM
•
Pump
temperatures
•
Pump current A
•
Pump power W
•
K1
condition
•
K2
condition
The controller is factory-set to HOST.
•
Confirm the selection by pressing the PUMP
CURRENT push-button.
•
In this way you enter into an operating phase
named "Monitor Relay Programming" described in
the following paragraph.
In case of RS 485 interface, the controller will display:
•
Press the PUMP CURRENT push-button to enter
the value and if the serial RS 485 option is installed,
the display shows:
S E R
I
A L
A D D R E S S
X X
where:
XX
= is the RS 485 address range [00-31]
The controller factory set is 00.
•
Select the desired value by pressing the CYCLE
NUMBER
S E R
I
A L
A D D R E S S
0
0