ProStar 325 UV-Vis Detector Control Manual
-
7 -
03-914947-44:2
Option Window
Press the
button to enter a code which
will enable the Dual Wavelength mode in the instrument. Then
check or uncheck the Two Wavelength Option to use it in
Galaxie.
Select the
Cell Type
according to the one physically installed in
the detector, four cell types are selectable: the 9x0, 9x1, 4x0
and 4x0.15.
If a prep flow cell is selected, the user will be able to enter a
Cell Ratio
. The ratio is used for extended range operation. The
ratio for each flow cell is determined at the factory and is
included with the flow cell. The detector will always operate in
extended range mode with a prep flow cell.