ProStar 325 UV-Vis Detector Control Manual
-
15 -
03-914947-44:2
applied to the analog outputs. It affects only the recorder and
data system analog outputs from the detector module, and does
not affect the data sent to the Galaxie Chromatography Data
System.
Enter the time program for the
Wavelength
, the
Attenuation,
and
the
Autozero
parameters.
The attenuation parameter switches the detector's output
attenuation to one of the eleven available settings. Attenuation is
measured in Absorbance Units Full Scale (AUFS). This affects
only the recorder’s output.
If
Yes
is selected in the
Autozero
field, the detector’s output is
automatically zeroed. This affects both the chromatogram display
and the data acquired for permanent storage.
Each step of the program includes time, wavelength, attenuation,
and autozero values.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
If
Dual Wavelength Mode
is checked extra columns are added in
the table which will allow the user to enter another Wavelength,
Attenuation and an Autozero value.