Reference
Page 74
User Manual Micro-GC
Varian, Inc.
Error
Class
©)
Error
number
(NNN)
Event/Error code
Description
Action needed
0
29
GC module changed
Changing a channel (controller or module) and
restarting the instrument
NA
0
30
TCD Gain calibrated
End TCD Gain calibration
NA
0
31
TCD Offset calibrated
End of Offset calibration
NA
0
32
Null String
Not used
NA
0
33
ADC reading out of range
Analog Digital Control out of range
NA
0
34
EDS Analytical Module incorrect
Electronic Data Sheet Analytical Module
incorrect
NA
0
35
EDS Config checksum incorrect
Electronic Data Sheet Configuration checksum
incorrect
NA
0 36
EDS Logbook checksum
incorrect
Electronic Data Sheet Logbook checksum
incorrect
NA
0 37
EDS Protected checksum
incorrect
Electronic Data Sheet Protected checksum
incorrect.
NA
0 38
EDS C.C. Config checksum
incorrect
Electronic Data Sheet Channel Control
checksum incorrect.
NA
0 39
EDS C.C. Logbook checksum
incorrect
Electronic Data Sheet Channel Control Logbook
checksum incorrect
NA
0 40
EDS C.C. Protected checksum
incorrect
Electronic Data Sheet Channel Control
Protected checksum incorrect
NA
0 41
EDS A.M. Config. checksum
incorrect
Electronic Data Sheet Analytical Module
Configuration checksum incorrect
NA
0 42
EDS A.M. Logbook checksum
incorrect
Electronic Data Sheet Analytical Module
Logbook checksum incorrect
NA
0 43
EDS A.M. Protected checksum
incorrect
Electronic Data Sheet Analytical Module
Protected checksum incorrect
NA
0
44
EDS Config SVER incorrect
Electronic Data Sheet Configuration Structure
Version incorrect
NA
0
45
EDS Protected SVER incorrect
Electronic Data Sheet Protected Structure
Version incorrect
NA
0 46
EDS C.C. Config SVER
incorrect
Electronic Data Sheet Channel Control Structure
Version incorrect
NA
0 47
EDS C.C. Protected SVER
incorrect
Electronic Data Sheet Channel Control
Protected Structure Version incorrect
NA
0 48
EDS A.M. Config SVER
incorrect
Electronic Data Sheet Analytical Module
Configuration incorrect
NA
0 49
EDS A.M. Protected SVER
incorrect
Electronic Data Sheet Analytical Module
Protected Structure Version incorrect
NA
0 50
Pressure Offset calibration
complete
Notification Pressure Offset calibration is
completed
NA
0 51
Pressure Offset calibration
Failed
Calibration offset out of range
0
52
Unable to store pressure offset
Pressure off set is out of valid range (range??)
2 53
Temperature sensor
disconnected
Temperature sensor not connect to instrument
Call Service
1
54
Not ready to start run
Issued by Safety Control Object in Hardware
domain. Bridge Call to GC domain (Reporting
Not Ready To Start Run Error)
Check Method
1
54
Stream selection failed
Stream selector (VICI) failed switching
Check Valve
1 55
Ambient pressure or
temperature alarm
Issued by Safety Control Object in Hardware
domain whenever ambient temperature has
passed a certain value.
1
56
Column cleaning
Instrument in column cleaning state
NA
1
57
Equilibrating temperature zones
Instrument stabilizing after column cleaning
Wait till Ready
60
DMD control to temperature
setpoint
Notification that DMD is warming up.
Wait for
completion