959D Leak Detector
viii
DRAFT
9/13/02
O-ring Care
When removing, checking, or replacing O-rings, keep in mind the following:
NOTE
Varian Vacuum Technologies recommends replacing all
O-rings during routine maintenance or during any maintenance
procedure requiring that O-rings be removed.
CAUTION
Remove O-rings carefully with your fingers. Do not use metal
tools for this task. Follow these guidelines to prevent scratching
of any sealing surfaces:
❑
Wipe all O-rings clean with a lint-free cloth before
installation to ensure that no foreign matter is present to
impair the seal.
❑
Do not use grease or use other substances on O-rings that
will come in contact with the spectrometer tube.
❑
Do not use alcohol, methanol, or other solvents on O-rings.
To do so causes deterioration and reduces the O-ring’s
ability to hold a vacuum.
❑
In allowable situations, apply a small amount of Apiezon
L
grease and wipe the O-rings “shiny” dry.
NOTE
Included in the Component and Spectrometer Tube Cleaning
Kit (p/n 670029096), VacuSolv is recommended for cleaning
the spectrometer tube components.
VacuSolv can also be used for fine cleaning of other parts in the
leak detector’s vacuum system, such as valves and fittings. No
rinsing steps or high-temperature drying is required following
cleaning with VacuSolv. Although appropriate precautions are
advised, VacuSolv is compatible with most materials and does
not contain toxic chemicals or CFCs (chlorofluorocarbons).
Содержание 959D
Страница 16: ...959D Leak Detector 1 2 D R A F T 9 1 3 0 2 Figure 1 1 Model 959D Leak Detector ...
Страница 28: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Страница 48: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Страница 68: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Страница 78: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Страница 81: ......