959D Leak Detector
3-15
3. Check that the ground slit plate is firmly seated in the ion source cavity.
If it is not, remove it and, using a screwdriver, bend the prongs out
slightly
and place the
ground slit plate into the cavity again.
WARNING
When handling the ion source, wear powder-free butyl or
polycarbonate gloves. If you handle the ion source with bare
hands, wash them thoroughly and especially before smoking or
eating.
4. Place the O-ring on the ion source as shown in Figure 3-15.
Figure 3-15
Replacing the Ion Source O-ring
When replacing the preamplifier and ion source, have the key of the center post line up
perpendicular to the spectrometer tube body. This alignment will happen by default for
some pieces. Replace the flange cap with the groove aligned with the two legs on either
side of the key.