959D Leak Detector
3-13
3. Use a Scotch-Brite
™
pad to polish away any discolored areas of the ground slit plate
(Figure 3-13).
Figure 3-13
Cleaning the Ground Slit Plate
CAUTION
The ground slit plate is very thin. Take care not to bend or
scratch it during cleaning.
4. Using a VacuSolv wipe, clean the ground slit plate, the flange cap, and wipe the cavity
and the outside of the spectrometer tube.
Let all the parts air dry. Take great care not to scratch any parts when wiping or when
setting onto a surface to dry.
5. Inspect all O-rings. Wipe with a lint-free cloth to be sure they are free of debris.
CAUTION
Remove O-rings carefully with your fingers. Do not use metal
tools for this task.
Follow these guidelines to prevent scratching of any sealing
surfaces:
❑
Wipe all O-rings clean with a lint-free cloth before
installation to assure that no foreign matter is present to
impair the seal.
❑
Do not use grease or use other substances on O-rings that
will come in contact with the spectrometer tube.
❑
Do not use alcohol, methanol, or other solvents on O-rings.
To do so causes deterioration and reduces the O-ring’s
ability to hold a vacuum.
❑
In allowable situations, apply a small amount of Apiezon
L
grease and wipe the O-rings “shiny” dry.
Содержание 959D
Страница 16: ...959D Leak Detector 1 2 D R A F T 9 1 3 0 2 Figure 1 1 Model 959D Leak Detector ...
Страница 28: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Страница 48: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Страница 68: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Страница 78: ...959D Leak Detector This page intentionally left blank D R A F T 9 1 3 0 2 ...
Страница 81: ......