03-914998-55:6
26 of 63
Figure 3.3
3.3 Dipole Adjustment
Go to Diagnostics. Check the RF Control check box and select the RF radio
button from the Monitoring tab as shown in Figure 3.4. Adjust the trapping field
dipole inductors on the Upper Manifold PWA through the holes in the Manifold
Electronics Enclosure Cover (as shown in Figure 3.5) so the Entrance and Exit
Endcap Trapping Field Dipole values equal 100%.