03-914998-55:6
36 of 63
Figure 5.1
5.2 OFN Acquisition
Use the Inject Single Sample option to make a 1 µL manual injection of test
standard 03-931126-01 (split style injection) or a 10:1 diluted test standard
(splitless pressure pulse style injection). Use a lower air gap to avoid sample loss
through the tip of the needle. Ensure the syringe plunger is not forced out of the
syringe barrel if a Pressure Pulse is used in the method.
5.3 Library Search and Signal to Noise Calculation
After acquiring the data file, calculate the S/N and run a library search. This
opportunity could be used to demonstrate data analysis to the customer. The
.sms data file can be loaded into Mass Spectra Data Review (MSDR) using the
quick load feature shown in Figure 5.2.
Содержание 4000 GCMS
Страница 4: ...03 914998 55 6 4 of 63 This page intentionally left blank...
Страница 10: ...03 914998 55 6 10 of 63 This page intentionally left blank...
Страница 13: ...03 914998 55 6 13 of 63 Select Install the software automatically Recommended and click Next Figure 2 2...
Страница 14: ...03 914998 55 6 14 of 63 Click Finish to close the wizard Figure 2 3...
Страница 17: ...03 914998 55 6 17 of 63 Figure 2 6...
Страница 18: ...03 914998 55 6 18 of 63 Figure 2 7...
Страница 25: ...03 914998 55 6 25 of 63 Figure 3 2...
Страница 27: ...03 914998 55 6 27 of 63 Figure 3 4 Figure 3 5...
Страница 30: ...03 914998 55 6 30 of 63 Figure 3 8...
Страница 34: ...03 914998 55 6 34 of 63 This page intentionally left blank...
Страница 40: ...03 914998 55 6 40 of 63 Figure 5 6...
Страница 41: ...03 914998 55 6 41 of 63 Figure 5 7...
Страница 42: ...03 914998 55 6 42 of 63 Figure 5 8...
Страница 50: ...03 914998 55 6 50 of 63 This page intentionally left blank...
Страница 58: ...03 914998 55 6 58 of 63 Figure 8 1...