1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
56
www.vareximaging.com
U
SER
M
ANUAL
13.3
Electromagnetic Immunity
Table 26
ESD, Transient/Burst, Surge, Voltage Variation, Magnetic Fields
Immunity Test
IEC 60601-1-2 Test
Level
Compliance
Electromagnetic
Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV
contact discharge
± 2, 4, 8, 15 kV
air discharge
± 8 kV
contact discharge
± 2, 4, 8, 15 kV
air discharge
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV AC Mains
±2 kV I/O Lines
±2 kV AC Mains
±2 kV I/O Lines
Mains power quality should
be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
±0.5 kV, ±1 kV
Line to Line
±0.5 kV, ±1 kV, ±2 kV
Line to Ground
±0.5 kV, ±1 kV
Line to Line
±0.5 kV, ±1 kV, ±2 kV
Line to Ground
Mains power quality should
be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Voltage dips:
0% UT
(100% dip in UT)
for 0,5 cycle
at 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% UT
(100% dip in UT)
for 1 cycle at 0°
70% UT
(30% dip in UT)
for 25/30 cycles
at 0°
Voltage Interruptions:
0% UT
(100% dip in UT)
for 250/300 cycle
Voltage dips:
0% UT
(100% dip in UT)
for 0,5 cycle
at 0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°
0% UT
(100% dip in UT)
for 1 cycle at 0°
70% UT
(30% dip in UT)
for 25/30 cycles
at 0°
Voltage Interruptions:
0% UT
(100% dip in UT)
for 250/300 cycle
Mains power quality should
be that of a typical commercial
or hospital environment. If the
user of the detector requires
continued operation during
power mains interruptions, it
is recommended that the
system be powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
Magnetic field should be that
of a typical location in a
typical commercial or hospital
environment.