www.vareximaging.com
55
1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
U
SER
M
ANUAL
13.2
Electromagnetic Emissions
Table 25
Radiated/Conducted Emissions, Harmonics, Voltage, Fluctuations, & Flicker
Emissions test
IEC 60601-1-2
test level
Compliance
Electromagnetic environment
RF conducted emissions
EN55011/CISPR11
Group 1, Class B,
150 kHz – 30 MHz
Group 1, Class B,
150 kHz – 30 MHz
The detector uses RF energy for
its internal function. Nearby
electronic equipment may be
affected.
RF radiated emissions
EN55011/CISPR 11
Group 1, Class B,
30 MHz – 1 GHz
Group 1, Class B,
30 MHz – 1 GHz
The detector uses RF energy for
its internal function. Nearby
electronic equipment may be
affected.
Harmonic emissions
EN/IEC 61000-3-2
Class A
Class A
The detector is suitable for use in
all establishments other than
domestic and those directly
connected to the low voltage
power supply network that
supplies buildings used for
domestic purposes.
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Limits defined
clause 5 of standard
The detector is suitable for use in
all establishments other than
domestic and those directly
connected to the low voltage
power supply network that
supplies buildings used for
domestic purposes.