1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
42
www.vareximaging.com
U
SER
M
ANUAL
Figure 43
Gain Calibration Flat Field Accumulation
7
Once flat field accumulation is complete, terminate the
X-ray beam
.
8
Wait
20 seconds
, then click
Continue
.
Figure 44
Dark Field Accumulation Begin