4343W/4336W-G5/2530W-G5 X-ray Detectors
Table 21
4336W-G5 and 2530W-G5 ESD, Transient/Burst, Surge, Voltage Variation, Magnetic
Fields (Continued)
Immunity test
IEC 60601-1-2
test level
Compliance
Electromagnetic environment
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Magnetic field should be that of a typical
location in a typical professional healthcare
environment.
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6V/m (in ISM bands
between 0.15MHz and
80MHz)
80% AM (at 1kHz)
3 Vrms
150 kHz to 80
MHz6V/m (in ISM
bands between
0.15MHz and 80MHz)
80% AM (at 1kHz)
-
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
3 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
E
=
6
--
P
d
Where
P
is the maximum power in
W
,
d
is
the minimum separation distance in
m
and
E
is the Immunity Test Level in V/m.
If the X-ray detector complies with
Immunity Test Levels for this test, the 30cm
minimum separation distance (in 5.2.1.1 f)
may be replaced with minimum separation
distances calculated from the higher
Immunity Test Levels.
Table 22
4336W-G5 and 2530W-G5 Test Specs for Enclosure Port Immunity to RF Wireless
Communications Equipment
Test
Frequency
Band
Service
Modulation
Max
Power
Distance
Immunity
Test Level
385
380-390
TETRA 400
Pulse Modulation
b) 18 Hz
1.8
0.3
27
450
430-470
GMRS 460, FRS
460
FM c) +/- 5 kHz
deviation 1 kHz sine
2
0.3
28
710
745
780
704-787
LTE BAND 13,
17
Pulse modulation
b) 217 Hz
0.2
0.3
9
810
870
930
800-960
GSM 1800;
TETRA 800;
iDEN 820;
CDMA 850;
LTE Band 5
Pulse modulation
b) 18 Hz
0.3
28