4343W/4336W-G5/2530W-G5 X-ray Detectors
14.1.2 Cleaning the X-ray Detector
To clean the X-ray detector:
1
Remove the battery from the X-ray detector.
2
Locate and read the cleaning instructions specified on the product label. Follow the product
instructions for cleaning.
Note
If you are using a disinfectant other than those specified, we recommend
you consult a specialist for the procedure for disinfection.
3
Dispense wipe from container.
4
Visually inspect the device. If contamination is present, use a wipe to remove the
contamination from the device surface. Use multiple wipes, as applicable, to remove the
visible contamination.
5
Take care to remove contamination from seams, joints, and other difficult-to-reach areas.
6
Dispose of each wipe used for cleaning.
7
If visible residue from the cleaning product is evident after air-drying, remove the residue
with a general Isopropyl Alcohol wipe, and let air-dry.
14.1.3 Disinfecting the X-ray Detector
To disinfect the X-ray detector:
1
Remove the battery from the X-ray detector.
2
Locate and read the cleaning instructions specified on the product label. Follow the product
instructions for cleaning.
Note
If you are using a disinfectant other than those specified, we recommend
you consult a specialist for the procedure for disinfection.
3
Wipe the entire surface of the target areas of the system until they are visibly wet with the
chemical solution. Target areas include any surface that was cleaned and/or any other
surfaces potentially contaminated during system use.
4
Take care to wet seams, joints, and other difficult-to-reach areas.
5
Ensure that the surfaces remain visibly wet for the maximum necessary disinfection time
specified on the product label.
a
If a disinfection time is not specified on a chemical label for the concentration used,
ensure that the surfaces remain visibly wet for a minimum of
10 minutes
.