43
4343W/4336W-G5/2530W-G5 X-ray Detectors
To reduce the effects of noise, the average of each pixel in the flat field image is calculated by
accumulating a number of frames into an internal memory buffer, then dividing the sum of each
pixel by the number of frames acquired.
Note
Gain calibration requires X-ray dose. Certain precautions must be taken by
the human operator.
The number of calibration frames used during Gain and Offset calibrations can be adjusted
under the Mode Settings pull down menu.
Note
We recommend accumulating 32 frames for gain calibration and 8 frames for
offset calibration for optimal image quality.
Note
Gain calibration should be performed at least once per year.
The actual number of calibration frames used is determined by the system integrator depending
upon their specific performance requirements.
Note
ViVA provides the convenience of automatically running Offset Calibration
as part of the gain calibration process.
However, API driven gain calibrations do not automatically run Offset
Calibration. OEM Applications should be sure to run Offset Calibration
prior to gain calibration.